■BROADBAND NMR POWER AMPLIFIER N146-554AA |
This device is designed to act as a high performance power amplifier for Nuclear Magnetic Resonance (NMR). It is an all solid-state high-speed pulse amplifier. |
Low noise is achieved by adopting a bias gate system. |
It can be used to connect CONTROL I/O connector with AMP CONTROL connector of our PROT series as our PROT system. |
|
Frequency range | 1MHz - 400MHz |
Output power | 300W/50Ω |
Input level | 0dBm |
Power gain | 55dB typ |
Gain flatness | 4dB |
Duty | Less than 10% |
Pulse width | Less than 1ms |
Input connector | BNC-FEMALE |
Input/Output impedance | 50Ω |
Output connector | N-FEMALE |
Gate pulse input connector (TX1) | BNC-FEMALE |
Preceding gate input connector (TX2) | BNC-FEMALE (10µs ahead TX1) |
Input monitor | RF output: -20dB |
Output monitor | FWD/BWD wave (Digital meter) FWD/BWD wave (RF output: -60dB) |
Protective Function | Excessive reflection (BWD: >75W) Abnormal temperature (TEMP: Heat sink >60℃) DC supply failure (PS:DC voltage monitoring) EXcessive pulse width (Pulse width: >11ms) |
Shape | 19 inches JIS rack case |
Size | W480xH149xD450 (mm) |
Weight | Less than 12kg |
Power supply | AC100V 50/60Hz, Less than 1kVA |
■BROADBAND NMR POWER AMPLIFIER N146-55ABA |
This device is designed to act as a high performance power amplifier for Nuclear Magnetic Resonance (NMR). It is an all solid-state high-speed pulse amplifier. |
Low noise is achieved by adopting a bias gate system. |
It can be used to connect CONTROL I/O connector with AMP CONTROL connector of our PROT series as our PROT system. |
|
Frequency range | 400MHz - 800MHz |
Output power | 300W/50Ω |
Input level | 0dBm |
Power gain | 55dB typ |
Gain flatness | 4dB |
Duty | Less than 10% |
Pulse width | Less than 1ms |
Input connector | BNC-FEMALE |
Input/Output impedance | 50Ω |
Output connector | N-FEMALE |
Gate pulse input connector (TX1) | BNC-FEMALE |
Preceding gate input connector (TX2) | BNC-FEMALE (10µs ahead TX1) |
Input monitor | RF output: -20dB |
Output monitor | FWD/BWD wave (Digital meter) FWD/BWD wave (RF output: -60dB) |
Protective Function | Excessive reflection (BWD: >75W) Abnormal temperature (TEMP: Heat sink >60℃) DC supply failure (PS:DC voltage monitoring) EXcessive pulse width (Pulse width: >11ms) |
Shape | 19 inches JIS rack case |
Size | W480xH299xD550 (mm) |
Weight | Less than 20kg |
Power supply | AC100V 50/60Hz, Less than 1kVA |